Skip to content
Thick Film Heaters · UL 499 / IEC 60335 / RoHS

Stainless Steel Substrate (thick film) Medical Fluid Warming Thick Film Plates

The temperature this substrate reaches, the watt density it carries, how it has to be clamped, and what it will not tolerate mechanically.

Stainless Steel Substrate (thick film) — Substrate to +500 °C; element system typically rated to +400 °C. A fired glass dielectric and resistor paste on 304 or 316 stainless. The heated surface is the pressure boundary itself, so there is no interface to conduct across. In this construction it carries 5 W/cm² in free air and 30 W/cm² bonded over its whole area to metal.

Insulation system
Stainless Steel Substrate (thick film)
Temperature range
Substrate to +500 °C; element system typically rated to +400 °C
Where it stops
Rigid, and the fired layers do not tolerate flexing or a dented substrate.
Continuous temperature limit
150 °C continuous
Watt density
5 W/cm² free air, 30 W/cm² bonded
Total thickness
1.5 mm to 3 mm

The temperature ceiling belongs to the insulation, not to the element: the resistance alloy would run far hotter, and it is the insulation around it that fails first. Figures are for the insulation system as built here; a part is quoted against the drawing, not against this table.

Medical Fluid Warming Thick Film Plates — the construction this guide is about

Stainless Steel Substrate (thick film): A fired glass dielectric and resistor paste on 304 or 316 stainless. The heated surface is the pressure boundary itself, so there is no interface to conduct across. Its working range is Substrate to +500 °C; element system typically rated to +400 °C, and that range is the property everything else on the part is designed around, because the element will always outlast the insulation around it.

Stainless Steel Substrate (thick film) as an Insulation System

A fired glass dielectric and resistor paste on 304 or 316 stainless. The heated surface is the pressure boundary itself, so there is no interface to conduct across.

Working rangeSubstrate to +500 °C; element system typically rated to +400 °C
Total thickness in this construction1.5 mm to 3 mm
Dielectric withstand2500 V AC, 1 minute
Insulation resistancegreater than 100 Mu03a9 at 500 V DC
Maximum size, one piece120 x 200 mm

Here the insulation is fired, not laminated, and the substrate is a structural part of the heater rather than a jacket around it. That is what allows the temperature capability and the watt density; it is also what makes the part rigid, and rigid parts have to be clamped flat rather than wrapped.

The Temperature Ceiling, and What Sets It

Working range: Substrate to +500 °C; element system typically rated to +400 °C. The ceiling belongs to the insulation. Nichrome will run red; stainless steel substrate (thick film) will not, and it is the material around the element that decides what the part is rated at.

This construction is rated 150 °C continuous. That is a continuous figure at the element, not an ambient and not a surface temperature measured somewhere convenient — and it is not a figure that derating extends downwards from something higher. Past the ceiling the insulation degrades, the dielectric strength falls with it, and the failure is electrical rather than thermal.

The boundaries across this catalogue are all material boundaries: PET stops at 120 °C, filled epoxy at 150 °C, silicone rubber and polyimide at 200 °C, and above that only fired and inorganic systems — thick film on stainless or alumina, and mica — are candidates at all.

What Attacks Stainless Steel Substrate (thick film)

Rigid, and the fired layers do not tolerate flexing or a dented substrate.

This construction is wetted in normal service, so the barrier is a design item and not a material property: the seal, the lead exit and the ingress rating are on the drawing, and the dry-run case is what the protection is sized for.

What a Medical Fluid Warming Thick Film Plates Is

Low mass, fast, closely controlled plates for blood and infusion warming, dialysis and analyser reagent lines, built under an ISO 13485 system.

Dielectric and resistor pastes screen printed onto stainless steel or alumina and fired at around 850 °C, then laser trimmed to resistance. There is no separate element and no interface: the heated surface is the substrate itself. In a fluid module the far face is the wetted wall, so the water carries the heat away as fast as the paste produces it — which is what allows watt densities an order of magnitude above any flexible construction, and what makes dry-run protection mandatory rather than advisable.

The heat leaves through a clamped joint, and a clamped joint is only as good as the pressure and the flatness behind it. A thermal interface material or a thin film of thermal grease fills the gaps that machining leaves; without one, contact happens at the high spots and the element runs hotter than the plate it is bolted to.

Elementfired resistor paste, laser trimmed
Insulation systemfired glass dielectric on the substrate
Total thickness1.5 mm to 3 mm
Maximum size, one piece120 x 200 mm
Continuous working temperature150 °C

The continuous limit of 150 °C is set by fired glass dielectric on the substrate. The element alloy would run far hotter; it is never the element that decides.

What the Insulation Changes About the Heater

A fired system removes the interface between element and heated surface, which is what allows the watt density to be an order of magnitude above any flexible construction. What it takes away is compliance: the part is the shape it was made, it has to be clamped flat, and the assembly around it has to accommodate its expansion rather than the other way round.

Attachment and Thermal Contact

The part is clamped, not bonded. What matters at the joint is flatness, contact pressure and a thermal interface material to fill what machining leaves behind — a dry joint between two nominally flat surfaces touches on perhaps a few percent of its area.

Clamp evenly and to a torque, not to feel. A ceramic substrate in particular survives 400 °C far better than it survives one bolt pulled down harder than the other three.

From the capability table: Bolted, brazed or welded into the assembly; flatness of the mating face is critical.

The joint between the heated surface and the load carries every watt the element makes. Fill it: an interface pad, a thin film of thermal grease or a bonded assembly. Two dry surfaces bolted together are a thermal resistance, and that resistance appears as element temperature.

Watt Density: the Two Figures

This construction is rated at 5 W/cm² in free air and 30 W/cm² bonded over its whole area to metal. The second figure is 6 times the first, and the difference is not a safety margin — it is the heat sink. On a 100 x 150 mm outline, which is 150 cm², that is about 75 W hanging in air against about 45 W bonded down.

Specify at the figure that matches how the heater will actually run. A heater ordered at the bonded density and then switched on before it is attached — on the bench, during a test, after it has lifted off in service — reaches its insulation limit in minutes, because there is nowhere for the heat to go but through the insulation.

Thick film reaches these densities because the heated surface is the substrate — there is no interface between element and load to conduct across. That is also why dry-run protection is not optional: with the fluid gone, so is the entire thermal path, and the fired layers reach their limit in seconds rather than minutes.

Control and Protection

This heater has no way of limiting its own temperature. Resistance heating draws roughly the same power whatever the temperature, so with the load removed, the fluid gone, the fan stopped or the sensor open circuit, the element keeps producing watts until something stops it.

The design must therefore contain a control device and, separately, a protective one: a controller with a sensor to hold the set point, and a bimetal thermostat or a one-shot thermal cutoff wired in series with the element to open the circuit if the loop ever fails. The two are not interchangeable — a controller that has lost its sensor cannot protect anything, which is exactly the fault the cutoff exists for.

On a fluid heater the limit device belongs on the substrate, close to the resistor, not downstream in the water. Downstream is where the temperature has already gone wrong; the substrate is where it goes wrong first, and at these watt densities the interval between the two is short.

Sensors can be built into the part: NTC thermistor, PT100 or PT1000 RTD, or a K-type thermocouple, fitted under the same insulation and tested with the heater as one assembly. A sensor laminated into the heater reads the heater; a sensor on the load reads the load. Both are legitimate, and the drawing has to say which is wanted.

T

h

How This Construction Fails

  • Run at the bonded watt density in free air. The insulation reaches its limit within minutes and the element burns through where the local density is highest. Specify at 30 W/cm² only where the heater is genuinely bonded to metal over its whole area; unbonded, the figure is 5 W/cm².
  • Loss of the control loop. An open-circuit sensor, a welded relay contact or a controller in manual leaves the element at full power with nothing watching it — the case the series thermal cutoff exists for, and the reason it must not be omitted to save a part.
  • Dry running. With the fluid gone the entire thermal path is gone, and at these densities the fired layers and the trim are destroyed in seconds. Flow or level interlocks, not a downstream thermostat, are what prevent it.
  • Scale on the wetted face. A layer of scale is a thermal resistance in exactly the wrong place: the substrate temperature rises to push the same power through it, and the failure looks like an overpowered heater rather than a fouled one.

Leads and Terminations

Termination is specified on the drawing: lead length, gauge, insulation and the connector or tag at the end of it.

Connector and termination options: Welded tabs, crimp terminals or soldered leads depending on temperature. A keyed housing removes a soldering operation from the customer's line and removes the possibility of the assembly being connected backwards; on a heater with a separate sensor circuit it also stops the sensor pair being mistaken for the power pair.

The lead exit is a thermal discontinuity as well as an electrical one. The joint between a fine element and a stranded lead is the hottest point of many heaters, so it is kept out of the highest watt density zone, strain relieved, and — where the exit sits in a hot area — insulated in PTFE rather than in silicone.

How to Specify One

This is a made-to-drawing part, so there is no catalogue number to quote. A quotation is built from the following, and the more of it that arrives with the enquiry the fewer rounds it takes:

  • An outline drawing, or a sketch with dimensions: overall size, holes, cut-outs, radii and the lead exit position.
  • The supply voltage, and whether it is AC or DC. The element is wound for one voltage; it is not a switchable range.
  • The power required, or — better — the temperature to be reached and the mass or flow it has to be reached in. Given the second, the wattage can be calculated instead of guessed.
  • What the heater is attached to, and how. This is what decides which watt density figure applies, and it is the question a quotation cannot proceed without.
  • The ambient, and the lowest temperature the assembly must start from. A heater sized for a 20 °C start is undersized for a -30 °C one.
  • Duty cycle: continuous, intermittent, or a warm-up to a deadline. Time to temperature is a specification, not a by-product.
  • Sensor and protection: which sensor, where it goes, and which limit device is fitted in series with the element.
  • The certification the end product is built to, named on the enquiry rather than after approval.
  • Quantity and target price, because the construction that is right for fifty pieces is often not the one that is right for fifty thousand.
  • The fluid, its flow rate and its inlet temperature, and how a dry-run condition is detected in your system.

Lead time: Samples 7-10 days from approved drawing; production 15-20 days.

Medical Fluid Warming Thick Film Plates — Capability Table

These are process limits for the construction, not the specification of one stocked item. Every line is what the factory can hold; what your part holds is fixed on your drawing.

ParameterSpecificationLimit / note
ConstructionGlass dielectric and resistor pastes screen printed and fired onto the substrate, then laser trimmed to resistanceFired, bonded or encapsulated as a single assembly
Substrate316L stainless or aluminaThe substrate carries the heat and the mechanical load
Thickness1.5 mm to 3 mmSubstrate plus element system
Maximum size, one piece120 x 200 mmLimited by the firing or curing fixture
Dimensional toleranceu00b10.2 mm on the substrate, u00b10.1 mm on printed featuresOn the machined substrate
Supply voltage12 V DC to 400 V ACCircuit is designed for the ordered voltage
Resistance toleranceu00b110% standard, u00b15% selectedThick film resistors are laser trimmed where a tighter band is required
Watt density5 W/cmu00b2 typical, 30 W/cmu00b2 maximum in a wetted applicationAchievable because the heat is generated in the surface being cooled
Maximum working temperature+150 u00b0C at the element systemOf the element system; the substrate itself tolerates more
Dielectric strength2500 V AC, 1 minuteOne minute, circuit to substrate
Insulation resistancegreater than 100 Mu03a9 at 500 V DCAt 500 V DC
TerminationsWelded tabs, crimp terminals or soldered leads depending on temperatureSoldered, welded or crimped as the temperature allows
MountingBolted, brazed or welded into the assembly; flatness of the mating face is criticalFlatness of the mating face matters more than clamping force
Sensor and control optionsPrinted PTC or NTC track, or a bonded thermocoupleIntegrated into the same firing or bonded to the substrate
Certifications availableUL 499, IEC/EN 60335, RoHS, REACH, CEConstruction level
Control accuracyu00b10.5 u00b0C at the plate with a two-sensor loopOne sensor controls, the second is an independent over-temperature channel
RedundancyIndependent hardware cutoff separate from the software loopRequired by IEC 60601 for anything warming fluid that reaches a patient

Capability figures are this factory's process limits for fired thick film construction. Watt densities far above any flexible heater are achievable because the resistor is printed directly onto the surface being cooled — there is no interface and no adhesive in the heat path.

Outline & Specification Builder

Set the outline, the size, the voltage and the power you have in mind. The drawing redraws to match, and the panel works out the heated area, the watt density and the current — then checks that watt density against what this construction survives in free air and bonded to a heat sink. Those two limits are different by several times, and it is the difference that decides whether a heater lasts.

Fills in the form below; attach your outline drawing or DXF there.

Build the Drawing

The panel above checks whether a size and a wattage can be built. This is the drawing itself: set the outline, route the heating track, place the lead wires, terminals, connector and temperature sensor, working in metal-clad thick film, and turn the model to look at it. When it is right, export the 2D drawing as a PDF or a DXF and send it with your enquiry — a drawing is what a quotation for medical fluid warming thick film plates is actually written from.

HEATING FILM DRAWING
Aluminium / Metal-Clad Thick Film Heater

Draw the outline, route the heating track, place the lead exits, terminals and sensor, dimension it, then export a PDF or a DXF and send it with your enquiry. Nothing here is a stock part and nothing you draw commits you to anything.

CMD

Want to compare constructions before you draw? Open the full builder, where every film family is selectable in one place.

Our facility

The numbers come off our own machines

Every unit is measured for resistance, held at its dielectric test voltage for a minute, and measured again for insulation resistance. Sample units from each lot are run under power and imaged. That is why the values on this page are quotable rather than indicative.

Insulation film fed from the roll
Vulcanising press closing on a lay-up
Heater outlines being die cut
Finished heaters batched for dispatch

Request a Quote

The form below follows the current technical drawing configuration. Change the drawing values above and the request details update automatically.

Request a Quotation

The part

If you have an outline drawing or a DXF, say so in the notes and we will reply with an address to send it to — a drawing turns an estimate into a quotation.

Business Information

Dimensions & Specifications (UL 499 / IEC 60335 / RoHS)

ParameterSpecificationLimit / note
ConstructionGlass dielectric and resistor pastes screen printed and fired onto the substrate, then laser trimmed to resistanceFired, bonded or encapsulated as a single assembly
Substrate316L stainless or aluminaThe substrate carries the heat and the mechanical load
Thickness1.5 mm to 3 mmSubstrate plus element system
Maximum size, one piece120 x 200 mmLimited by the firing or curing fixture
Dimensional toleranceu00b10.2 mm on the substrate, u00b10.1 mm on printed featuresOn the machined substrate
Supply voltage12 V DC to 400 V ACCircuit is designed for the ordered voltage
Resistance toleranceu00b110% standard, u00b15% selectedThick film resistors are laser trimmed where a tighter band is required
Watt density5 W/cmu00b2 typical, 30 W/cmu00b2 maximum in a wetted applicationAchievable because the heat is generated in the surface being cooled
Maximum working temperature+150 u00b0C at the element systemOf the element system; the substrate itself tolerates more
Dielectric strength2500 V AC, 1 minuteOne minute, circuit to substrate
Insulation resistancegreater than 100 Mu03a9 at 500 V DCAt 500 V DC
TerminationsWelded tabs, crimp terminals or soldered leads depending on temperatureSoldered, welded or crimped as the temperature allows
MountingBolted, brazed or welded into the assembly; flatness of the mating face is criticalFlatness of the mating face matters more than clamping force
Sensor and control optionsPrinted PTC or NTC track, or a bonded thermocoupleIntegrated into the same firing or bonded to the substrate
Certifications availableUL 499, IEC/EN 60335, RoHS, REACH, CEConstruction level
Control accuracyu00b10.5 u00b0C at the plate with a two-sensor loopOne sensor controls, the second is an independent over-temperature channel
RedundancyIndependent hardware cutoff separate from the software loopRequired by IEC 60601 for anything warming fluid that reaches a patient

Video: Stainless Steel Substrate (thick film) Medical Fluid Warming Thick Film Plates

Third-party video, selected automatically for this topic. It is not produced by Heating Film and is not a specification source — use the tables on this page for dimensional data.

Frequently Asked Questions

What is the maximum temperature of medical fluid warming thick film plates?

150 °C continuous, measured at the element. The limit belongs to fired glass dielectric on the substrate and not to the element alloy, so it is not a figure that can be extended by derating the power.

Does this heater need a thermostat?

Yes. It has no mechanism for limiting its own temperature — with the load removed or the control loop failed it keeps producing the same power — so the design needs a controller with a sensor to hold the set point and, separately, a bimetal thermostat or a one-shot thermal cutoff in series with the element as protection.

Is there a part number I can order?

No — these are made to your drawing, so a part number would describe something that was never built. A quotation needs the outline, the supply voltage, the power or the target temperature, what the heater is bonded to, the ambient and the duty cycle. The table on this page is what the factory can hold, not a stock list.

How large can one piece be?

120 x 200 mm in a single piece. Larger areas are supplied as tiled zones with separate circuits, which is usually better engineering anyway: separate zones can be controlled separately, and a large single circuit has to carry the whole current at one lead exit.

How do I request a quotation?

Send the outline drawing with the supply voltage, the power or the target temperature, what the heater is bonded to, the ambient and the duty cycle. We respond within one business day.