ISO 13485 Medical Device Polyimide Heaters

What ISO 13485 fixes, what it deliberately leaves to another document, and how it is evidenced on a heater built to a customer drawing.

The quality management system standard for medical devices, and the one a heater destined for a diagnostic instrument or a patient-warming device is built under.

Standard
ISO 13485
What it does not cover
Not a device approval; the device manufacturer still carries the regulatory submission.
Construction
Medical Device Polyimide Heaters
Built and tested to
UL 499 / IEC 60335 / RoHS
Insulation
Polyimide film
Continuous temperature limit
200 °C continuous

A construction is built and tested to a standard. A heater made to your outline has no certified part number of its own, so the approval sits with the end product and with the file you hold — what this factory supplies is the construction, the materials and the test records that let that file be assembled.

Medical Device Polyimide Heaters — the construction this guide is about

A standard is a set of tests, not a property of a shape. ISO 13485 is one of the documents medical device polyimide heaters are built to; the sections below say what it fixes, what it does not, and what has to be named on the enquiry for the approval to be usable.

ISO 13485 — Scope

The quality management system standard for medical devices, and the one a heater destined for a diagnostic instrument or a patient-warming device is built under.

What ISO 13485 Does Not Cover

Not a device approval; the device manufacturer still carries the regulatory submission.

This is the division of labour to carry away from any heating standard: safety standards decide whether the part can hurt somebody, and say nothing about whether it holds a temperature, lasts five years or matches the last batch. Performance is a specification you write; safety is one the standard writes.

It is also a certification of the factory rather than of the part. It says how the part is controlled, documented and traced; it does not say the part is fit for your application, and it does not carry your regulatory submission for you.

How a Made-to-Drawing Heater Is Certified

A construction is built and tested to a standard. The insulation system, the creepage and clearance distances, the earthing, the lead anchorage, the temperature under fault and the marking are what the test covers — and all of those are properties of the construction rather than of the outline it is die-cut to.

So a custom outline does not have a certificate of its own, and no supplier can honestly hand you one. What is available is the construction file: the standard the build is tested to, the materials list with their own ratings, and the test reports. That is what an approvals engineer needs, and it is what the end product's own certification is built on.

The practical consequence: name ISO 13485 on the enquiry, not after approval. A construction chosen for cost and then asked to pass a standard afterwards usually has to be rebuilt, because the insulation thickness, the creepage distances and the lead anchorage are decided at design time and are not adjustable later.

From the capability table: UL 499 construction, IEC/EN 60335, UL 94 V-0, RoHS, REACH, CE, ISO 13485 QMS.

What a Medical Device Polyimide Heaters Is

PI heaters built under an ISO 13485 quality system for diagnostic instruments, sample warming, reagent cartridges and patient-contact assemblies.

A photo-etched copper or Inconel foil circuit laminated between polyimide films, 0.2 to 0.4 mm thick in total. It is the thinnest and fastest flexible heater made: almost no thermal mass of its own, so what it heats is the part rather than itself. It will wrap a cylinder; it will not take a compound curve, and a nick in a 0.05 mm film propagates from the edge, so outlines are drawn with radiused corners.

The heat leaves through the bonded face. Everything that is not in contact with the load radiates and convects from the back of the heater instead, which at these thicknesses is a poor path — so an air gap under part of the outline is not a small loss of efficiency, it is a hot spot at that spot.

Elementphoto-etched copper or Inconel foil
Insulation systempolyimide film
Total thickness0.25 mm
Maximum size, one piece300 x 500 mm (one piece)
Continuous working temperature200 °C

The continuous limit of 200 °C is set by polyimide film. The element alloy would run far hotter; it is never the element that decides.

The Other Standards This Construction Is Built To

A heater is rarely governed by one document. This construction is also built and tested to:

Safety, materials, substances and quality system are four separate questions with four separate documents behind them. A part can satisfy all four and still be the wrong heater for the duty, which is what the rest of this page is about.

Control and Protection

This heater has no way of limiting its own temperature. Resistance heating draws roughly the same power whatever the temperature, so with the load removed, the fluid gone, the fan stopped or the sensor open circuit, the element keeps producing watts until something stops it.

The design must therefore contain a control device and, separately, a protective one: a controller with a sensor to hold the set point, and a bimetal thermostat or a one-shot thermal cutoff wired in series with the element to open the circuit if the loop ever fails. The two are not interchangeable — a controller that has lost its sensor cannot protect anything, which is exactly the fault the cutoff exists for.

Sensors can be built into the part: NTC thermistor, PT100 or PT1000 RTD, or a K-type thermocouple, fitted under the same insulation and tested with the heater as one assembly. A sensor laminated into the heater reads the heater; a sensor on the load reads the load. Both are legitimate, and the drawing has to say which is wanted.

Watt Density: the Two Figures

This construction is rated at 0.5 W/cm² in free air and 5 W/cm² bonded over its whole area to metal. The second figure is 10 times the first, and the difference is not a safety margin — it is the heat sink. On a 100 x 150 mm outline, which is 150 cm², that is about 75 W hanging in air against about 75 W bonded down.

Specify at the figure that matches how the heater will actually run. A heater ordered at the bonded density and then switched on before it is attached — on the bench, during a test, after it has lifted off in service — reaches its insulation limit in minutes, because there is nowhere for the heat to go but through the insulation.

Leads and Terminations

Lead wire as supplied: PTFE or silicone insulated 24 AWG to 20 AWG, or an FFC tail etched as part of the circuit. Length, gauge, exit position and termination are drawing items — the exit position in particular, because it decides where the heater can be routed and it cannot be changed after the tool is cut.

The lead exit is a thermal discontinuity as well as an electrical one. The joint between a fine element and a stranded lead is the hottest point of many heaters, so it is kept out of the highest watt density zone, strain relieved, and — where the exit sits in a hot area — insulated in PTFE rather than in silicone.

On a 0.2 mm film the lead joint is thicker than the heater. Where the assembly has to stay flat, the leads are brought out to a tail or a flat flexible tab rather than soldered onto the heated area.

How to Specify One

This is a made-to-drawing part, so there is no catalogue number to quote. A quotation is built from the following, and the more of it that arrives with the enquiry the fewer rounds it takes:

  • An outline drawing, or a sketch with dimensions: overall size, holes, cut-outs, radii and the lead exit position.
  • The supply voltage, and whether it is AC or DC. The element is wound for one voltage; it is not a switchable range.
  • The power required, or — better — the temperature to be reached and the mass or flow it has to be reached in. Given the second, the wattage can be calculated instead of guessed.
  • What the heater is attached to, and how. This is what decides which watt density figure applies, and it is the question a quotation cannot proceed without.
  • The ambient, and the lowest temperature the assembly must start from. A heater sized for a 20 °C start is undersized for a -30 °C one.
  • Duty cycle: continuous, intermittent, or a warm-up to a deadline. Time to temperature is a specification, not a by-product.
  • Sensor and protection: which sensor, where it goes, and which limit device is fitted in series with the element.
  • The certification the end product is built to, named on the enquiry rather than after approval.
  • Quantity and target price, because the construction that is right for fifty pieces is often not the one that is right for fifty thousand.
  • Whether adhesive backing is wanted, and what surface it goes onto — a low-energy plastic needs a different adhesive from bare aluminium.

Lead time: Samples 7-10 days from approved drawing; production 15-20 days.

Medical Device Polyimide Heaters — Capability Table

These are process limits for the construction, not the specification of one stocked item. Every line is what the factory can hold; what your part holds is fixed on your drawing.

ParameterSpecificationLimit / note
ConstructionPhoto-etched copper or Inconel foil circuit laminated between polyimide films with a high temperature acrylic or FEP adhesiveDecided at quotation from the outline and the wattage
Insulation systemPolyimide film, 0.05 mm or 0.125 mm per faceThis is what sets the temperature limit, not the element
Total thickness0.25 mmNominal, before adhesive; add 0.2 mm for PSA backing
Minimum size10 x 10 mmBelow this the element cannot be routed with a safe edge margin
Maximum size, one piece300 x 500 mm (one piece)Larger areas are supplied as tiled zones with separate circuits
Dimensional toleranceu00b10.5 mm on dimensions up to 200 mmOn the die-cut outline; hole positions are held tighter
Supply voltage5 V DC to 240 V AC, wound to orderWound to the ordered voltage u2014 it is not a switchable range
Resistance toleranceu00b110% standard, u00b15% on requestMeasured at 20 u00b0C; hot resistance differs with the element alloy
Watt density, free air0.5 W/cmu00b2 typical, 0.8 W/cmu00b2 maximumNothing conducting heat away u2014 the honest figure for an unbonded heater
Watt density, bonded to heat sink5.0 W/cmu00b2 maximumOnly valid with full-area contact to metal; specifying at this figure and running in free air destroys the heater
Continuous working temperature-200 u00b0C to +200 u00b0C continuous (+220 u00b0C short term)Of the heater itself, measured at the element, not the ambient
Dielectric strength1500 V AC, 1 minute (2500 V AC on the 0.125 mm film build)One minute, element to any exposed surface, no breakdown
Insulation resistancegreater than 100 Mu03a9At 500 V DC, measured after the dielectric test
Lead wirePTFE or silicone insulated 24 AWG to 20 AWG, or an FFC tail etched as part of the circuitLength, gauge and termination to your drawing
AttachmentFree, 3M PSA backing, or bonded with a two-part thermally conductive adhesiveAdhesive-backed heaters need continuous contact pressure while curing
Sensor and control optionsNTC 10K B3950, PT100/PT1000, K-type thermocouple, KSD9700 bimetal thermostat, thermal cutoffFitted under the same insulation and tested as one assembly
Certifications availableUL 499 construction, IEC/EN 60335, UL 94 V-0, RoHS, REACH, CE, ISO 13485 QMSConstructions are built to these; the certificate covers the construction, not every outline
DocumentationMaterial declarations, lot traceability, and PPAP-equivalent submissionThe paperwork is the deliverable as much as the heater is
CleanlinessAssembled and packed in a controlled area; ionic contamination controlledSpecify the cleanliness requirement u2014 it changes the process, not just the packaging

Capability figures are this factory's process limits for etched-foil polyimide construction. The 200 °C continuous limit and the watt density ceiling (to about 7.75 W/cm² with heat sinking) are properties of the polyimide film and the adhesive system, and are published consistently across the industry.

Outline & Specification Builder

Set the outline, the size, the voltage and the power you have in mind. The drawing redraws to match, and the panel works out the heated area, the watt density and the current — then checks that watt density against what this construction survives in free air and bonded to a heat sink. Those two limits are different by several times, and it is the difference that decides whether a heater lasts.

Fills in the form below; attach your outline drawing or DXF there.

Build the Drawing

The panel above checks whether a size and a wattage can be built. This is the drawing itself: set the outline, route the heating track, place the lead wires, terminals, connector and temperature sensor, working in polyimide film, and turn the model to look at it. When it is right, export the 2D drawing as a PDF or a DXF and send it with your enquiry — a drawing is what a quotation for medical device polyimide heaters is actually written from.

HEATING FILM DRAWING
Polyimide (PI / Kapton) Film Heater

Draw the outline, route the heating track, place the lead exits, terminals and sensor, dimension it, then export a PDF or a DXF and send it with your enquiry. Nothing here is a stock part and nothing you draw commits you to anything.

CMD

Want to compare constructions before you draw? Open the full builder, where every film family is selectable in one place.

Our facility

Where these heaters are actually made

Circuit layout, element preparation, lamination and vulcanising, die cutting and lead attachment all happen under one roof, so a watt density limit and a temperature ceiling are set by processes we control and can evidence, not by whichever supplier had stock.

Insulation film fed from the roll
Vulcanising press closing on a lay-up
Heater outlines being die cut
Finished heaters batched for dispatch

Request a Quote

The form below follows the current technical drawing configuration. Change the drawing values above and the request details update automatically.

Request a Quotation

The part

If you have an outline drawing or a DXF, say so in the notes and we will reply with an address to send it to — a drawing turns an estimate into a quotation.

Business Information

Dimensions & Specifications (UL 499 / IEC 60335 / RoHS)

ParameterSpecificationLimit / note
ConstructionPhoto-etched copper or Inconel foil circuit laminated between polyimide films with a high temperature acrylic or FEP adhesiveDecided at quotation from the outline and the wattage
Insulation systemPolyimide film, 0.05 mm or 0.125 mm per faceThis is what sets the temperature limit, not the element
Total thickness0.25 mmNominal, before adhesive; add 0.2 mm for PSA backing
Minimum size10 x 10 mmBelow this the element cannot be routed with a safe edge margin
Maximum size, one piece300 x 500 mm (one piece)Larger areas are supplied as tiled zones with separate circuits
Dimensional toleranceu00b10.5 mm on dimensions up to 200 mmOn the die-cut outline; hole positions are held tighter
Supply voltage5 V DC to 240 V AC, wound to orderWound to the ordered voltage u2014 it is not a switchable range
Resistance toleranceu00b110% standard, u00b15% on requestMeasured at 20 u00b0C; hot resistance differs with the element alloy
Watt density, free air0.5 W/cmu00b2 typical, 0.8 W/cmu00b2 maximumNothing conducting heat away u2014 the honest figure for an unbonded heater
Watt density, bonded to heat sink5.0 W/cmu00b2 maximumOnly valid with full-area contact to metal; specifying at this figure and running in free air destroys the heater
Continuous working temperature-200 u00b0C to +200 u00b0C continuous (+220 u00b0C short term)Of the heater itself, measured at the element, not the ambient
Dielectric strength1500 V AC, 1 minute (2500 V AC on the 0.125 mm film build)One minute, element to any exposed surface, no breakdown
Insulation resistancegreater than 100 Mu03a9At 500 V DC, measured after the dielectric test
Lead wirePTFE or silicone insulated 24 AWG to 20 AWG, or an FFC tail etched as part of the circuitLength, gauge and termination to your drawing
AttachmentFree, 3M PSA backing, or bonded with a two-part thermally conductive adhesiveAdhesive-backed heaters need continuous contact pressure while curing
Sensor and control optionsNTC 10K B3950, PT100/PT1000, K-type thermocouple, KSD9700 bimetal thermostat, thermal cutoffFitted under the same insulation and tested as one assembly
Certifications availableUL 499 construction, IEC/EN 60335, UL 94 V-0, RoHS, REACH, CE, ISO 13485 QMSConstructions are built to these; the certificate covers the construction, not every outline
DocumentationMaterial declarations, lot traceability, and PPAP-equivalent submissionThe paperwork is the deliverable as much as the heater is
CleanlinessAssembled and packed in a controlled area; ionic contamination controlledSpecify the cleanliness requirement u2014 it changes the process, not just the packaging

Video: ISO 13485 Medical Device Polyimide Heaters

Third-party video, selected automatically for this topic. It is not produced by Heating Film and is not a specification source — use the tables on this page for dimensional data.

Frequently Asked Questions

What watt density can medical device polyimide heaters run at?

0.5 W/cm² with nothing conducting the heat away, and 5 W/cm² bonded over its whole area to metal. The two figures describe two different installations, not a range to choose from: a heater specified at 5 W/cm² and then run in free air reaches its insulation limit within minutes.

What is the maximum temperature of medical device polyimide heaters?

200 °C continuous, measured at the element. The limit belongs to polyimide film and not to the element alloy, so it is not a figure that can be extended by derating the power.

Does this heater need a thermostat?

Yes. It has no mechanism for limiting its own temperature — with the load removed or the control loop failed it keeps producing the same power — so the design needs a controller with a sensor to hold the set point and, separately, a bimetal thermostat or a one-shot thermal cutoff in series with the element as protection.

Is there a part number I can order?

No — these are made to your drawing, so a part number would describe something that was never built. A quotation needs the outline, the supply voltage, the power or the target temperature, what the heater is bonded to, the ambient and the duty cycle. The table on this page is what the factory can hold, not a stock list.

How do I request a quotation?

Send the outline drawing with the supply voltage, the power or the target temperature, what the heater is bonded to, the ambient and the duty cycle. We respond within one business day.